Publication details

Methodology of separation and ablation of corneal tissue using femtosecond and excimer laser

Authors

VLASÁK Ondřej VODIČKOVÁ Kristina ŠKORPÍKOVÁ Jiřina

Year of publication 2015
Type Appeared in Conference without Proceedings
MU Faculty or unit

Faculty of Medicine

Citation
Description In the ReLEx method electron microscopy images show greater irregularities in both optical and non-optical portion. The situation is better in method using the multiple energy. But the biggest unknown can be seen in the transition region between the non-optical and optical zones The transition zone is clearly visible in the images transition line. From the optical point of view, it may cause undesirable light scattering. Irregularity of the transition line is reduced using the higher separation energy. We therefore conclude that this method is very susceptible to the correct power setting.

You are running an old browser version. We recommend updating your browser to its latest version.