You are here:
Publication details
Fast and accurate compensation of signal offset for T2 mapping
Authors | |
---|---|
Year of publication | 2019 |
Type | Article in Periodical |
Magazine / Source | Magnetic Resonance Materials in Physics, Biology and Medicine |
MU Faculty or unit | |
Citation | |
Web | https://doi.org/10.1007/s10334-019-00737-3 |
Doi | http://dx.doi.org/10.1007/s10334-019-00737-3 |
Keywords | Algorithms; Least-squares analysis; Software |
Description | T2 maps are more vendor independent than other MRI protocols. Multi-echo spin-echo signal decays to a non-zero offset due to imperfect refocusing pulses and Rician noise, causing T2 overestimation by the vendor's 2-parameter algorithm. The accuracy of the T2 estimate is improved, if the non-zero offset is estimated as a third parameter. Three-parameter Levenberg--Marquardt (LM) T2 estimation takes several minutes to calculate, and it is sensitive to initial values. We aimed for a 3-parameter fitting algorithm that was comparably accurate, yet substantially faster. |