Remotely Detected NMR for the Characterization of Flow and Fast Chromatographic Separations Using Organic Polymer Monoliths
Authors | |
---|---|
Year of publication | 2011 |
Type | Article in Periodical |
Magazine / Source | Analytical chemistry |
Citation | |
Doi | http://dx.doi.org/10.1021/ac2010108 |
Description | An application of remotely detected magnetic resonance imaging is demonstrated for the characterization of flow and the detection of fast, small molecule separations within hypercrosslinked polymer monoliths. The hyper-cross-linked monoliths exhibited excellent ruggedness, with a transit time relative standard deviation of less than 2.1%, even after more than 300 column volumes were pumped through at high pressure and flow. Magnetic resonance imaging enabled high. resolution intensity and velocity-encoded images of mobile phase flow through the monolith. The images confirm that the presence of a polymer monolith within the capillary disrupts the parabolic laminar flow profile that is characteristic of mobile phase flow within an open tube. As a result, the mobile phase and analytes are equally distributed in the radial direction throughout the monolith. Also, in-line monitoring of chromatographic separations of small molecules at high flow rates is shown. The coupling of monolithic chromatography columns and NMR provides both real-time peak detection and chemical shift information for small aromatic molecules. These experiments demonstrate the unique power of magnetic resonance, both direct and remote, in studying chromatographic processes. |