You are here:
Publication details
Interpretive discrepancies caused by target values inter-batch variations in chemiluminescence immunoassay for SARS-CoV-2 IgM/IgG by MAGLUMI
Authors | |
---|---|
Year of publication | 2021 |
Type | Article in Periodical |
Magazine / Source | Journal of Medical Virology |
MU Faculty or unit | |
Citation | |
Web | https://onlinelibrary.wiley.com/doi/10.1002/jmv.26612 |
Doi | http://dx.doi.org/10.1002/jmv.26612 |
Keywords | anti- SARS- CoV- 2 IgG; anti- SARS- CoV- 2 IgM; CLIA; inter- assay batch variation; interpretative inconsistency |
Description | Plasma specimens from coronavirus disease 2019 patients were double-tested for anti-severe acute respiratory syndrome coronavirus 2 (SARS-CoV-2) antibodies by two different batches of MAGLUMI 2019-nCov immunoglobulin M/immunoglobulin G (IgM/IgG) assays to evaluate IgM/IgG levels, qualitative interpretation, antibody kinetics, and linearity of diluted specimen. Here we show that (i) high-level IgM specimens need to be diluted with negative human plasma but not kit diluents and (ii) measured anti-SARS-CoV-2 IgM/IgG concentrations are substantially higher with later marketed immunoassay batch leading to (iii) the change of qualitative interpretation (positive vs. negative) in 12.3% of specimens measured for IgM, (iv) the informative time-course pattern of antibody production only when data from different immunoassay batches are not combined. |