Publication details

Nebulizer particle size distribution measured by various methods

Authors

MIŠÍK Ondrej PEŠTÁLOVÁ Andrea BĚLKA Miloslav LÍZAL František

Year of publication 2022
Type Article in Periodical
Magazine / Source EDP Sciences, EPJ Web of Conferences
Citation
Web https://doi.org/10.1051/epjconf/202226401023
Doi http://dx.doi.org/10.1051/epjconf/202226401023
Attached files
Description Pari LC Sprint Star nebulizer is an inhaler device dedicated to delivering the liquid drug formulation into peripheral lung regions. The particle size is one of the critical parameters which determine the location of aerosol deposition within the human lungs. In this study, the particle size distribution of aerosol was measured by two different methods: Andersen Cascade Impactor (ACI) and Aerodynamic Particle Sizer (APS). Mass Median of Aerodynamic Distribution (MMAD), and Geometric Standard Deviation (GSD) were evaluated. MMAD gained by APS was approx. 7 µm, MMAD acquired by ACI was approx. 2.5 µm. According to the results, these two measuring methods are not very consistent, and the comparison of the particle size acquired by the different methods is not appropriate.

You are running an old browser version. We recommend updating your browser to its latest version.

More info