You are here:
Publication details
Light Output Function and Pulse-Shape Discrimination Capability of P-Terphenyl Organic Scintillator in Wide Energy Rangeorganic Scintillator in Wide Energy Range
Authors | |
---|---|
Year of publication | 2024 |
Type | Article in Periodical |
Citation | |
Web | https://papers.ssrn.com/sol3/papers.cfm?abstract_id=4628647 |
Keywords | Neutron-gamma analyser; P-Terphenyl scintillator; Light output; Pulse-shape discrimination (PSD); Neutron efficiency; Mono-energetic neutrons; Digital-signal processing |
Description | In this paper, we studied light-output properties, efficiency function as well as PSD capability of p-Terphenyl scintillator. We used mono-energetic neutron beams in the wide energy range from 2.5 to 19 MeV. The neutron-gamma spectrometric system, which we developed has been utilised for the apparatus spectra measurement, processing and evaluation. We carried out accelerator facility at PTB Braunschweig and Research Center Rez, Prague. A major contribution of this paper is the refinement of the existing data for this type of scintillator. Until now, no one has determined by measurement the light-output response of a p-Terphenyl scintillator over such a wide energy range. |