Publication details

Central venous pressure measurement during simultaneous infusion via central venous catheter.

Authors

ŠRÁMEK Vladimír NESTICKÁ K. TYLEČKOVÁ L ŠTĚTKA Pavel FENCL Marek

Year of publication 2004
Type Article in Periodical
Magazine / Source Intensive care medicine
MU Faculty or unit

Faculty of Medicine

Citation
Field Surgery incl. transplantology
Description CVP value changes significanty at pump infusion rates 100 ml/hr or usage of 3 perfusors at high rate (20ml/hr each). At lower rates the changes are minimal and fluid infusion need not be stopped to obtain correct CVP value.

You are running an old browser version. We recommend updating your browser to its latest version.