You are here:
Publication details
LIBS in Environmental Analysis
Authors | |
---|---|
Year of publication | 2005 |
Type | Article in Proceedings |
Conference | IV. International Conference on Inorganic Environmental Analysis |
MU Faculty or unit | |
Citation | |
Field | Analytic chemistry |
Keywords | Laser ablation; LIBS;Time resolved emission spectrometry; Soil |
Description | Possibility of copper and vanadium determination in soil samples by LIBS was studied. Linear calibration curves for both elements were obtained in concentration range of 10 - 200 mg/kg using pressed pellets with silver binder. Good parameters of calibration curves were obtained for delay time 20 /us for vanadium and 4.0 /us for copper. Limits of detection evaluated from calibration curves were found around 10 mg/kg. This experimental setup of LIBS is able to determine above-mentioned elements in soil samples in concentration range tens to hundreds mg/kg. |
Related projects: |