Publication details

Palladium biosensor

Authors

ADAM Vojtěch HANUŠTIAK Pavel KIZEK René KŘÍŽKOVÁ Soňa TRNKOVÁ Libuše BEKLOVÁ Miroslava ZEHNÁLEK Josef HORNA Aleš SURES Bernd

Year of publication 2007
Type Article in Periodical
Magazine / Source Electroanalysis
MU Faculty or unit

Faculty of Science

Citation
Field Electrochemistry
Keywords Palladium; biosensor; electrochemistry
Description In this paper we proposed a palladium(II) biosensor. The biosensor is based on determining of interactions between palladium(II) and metallothionein modified hanging mercury drop electrode by means of differential pulse voltammetry. We studied influence of two supporting electrolytes (potassium or sodium chloride) on the signals of the biosensor. Based on the results obtained we found potassium chloride (0.05 M) as the most suitable supporting electrolyte to determine palladium(II). The detection limit of the biosensor for palladium ions was evaluated as 100 nM with RSD about 10%. Moreover, we utilized the biosensor for measurement of the target molecule in the presence of human blood serum and human urine.
Related projects:

You are running an old browser version. We recommend updating your browser to its latest version.

More info