You are here:
Publication details
ET-AAS determination of aluminium in dialysis concentrates after continuous flow solvent extraction
Authors | |
---|---|
Year of publication | 2007 |
Type | Article in Periodical |
Magazine / Source | Journal of Pharmaceutical and Biomedical Analysis |
MU Faculty or unit | |
Citation | |
Field | Analytic chemistry |
Keywords | Aluminium; Continuous flow extraction (CFE); ET-AAS determination; Dialysis solutions |
Description | Conditions of a continuous flow extraction of aluminium acetylacetonate in acetylacetone and aluminium 8-hydroxyquinolinate into methylisobutylketone were studied. The analytical signal of the aluminium chelates present in the organic phase was measured at 309.3 nm using atomic absorption spectrometry with electrothermal atomization (ETAAS) at the flow rate ratio Faq/Forg = 3 for aqueous and organic phases. The acetylacetonate method was applied to the determination of aluminium in real dialysis concentrates. |
Related projects: |