Albumin-based or albumin-linked calibrators cause a positive bias in serum protein analysis by the biuret method

Authors

CHROMÝ Vratislav ŠVÁCHOVÁ Lenka NOVOSÁD Lukáš JARKOVSKÝ Jiří SEDLÁK Pavel HORÁK Petr DOBROVOLNÁ Hana HLAVÁČOVÁ Barbora

Year of publication 2009
Type Article in Periodical
Magazine / Source Clinical Chemistry and Laboratory Medicine
MU Faculty or unit

Faculty of Science

Citation
Field Biochemistry
Keywords calibration of toal serum protein by the biuret reaction
Description Albunin-based or linked calibrators are responsible for a 5% positive bias when used for serum protein determination by the biuret method