Albumin-based or albumin-linked calibrators cause a positive bias in serum protein analysis by the biuret method
| Authors | |
|---|---|
| Year of publication | 2009 |
| Type | Article in Periodical |
| Magazine / Source | Clinical Chemistry and Laboratory Medicine |
| MU Faculty or unit | |
| Citation | |
| Field | Biochemistry |
| Keywords | calibration of toal serum protein by the biuret reaction |
| Description | Albunin-based or linked calibrators are responsible for a 5% positive bias when used for serum protein determination by the biuret method |