Publication details

Piezoelectric and optical biosensors for detection of Bacillus atrophaeus

Investor logo
Authors

FARKA Zdeněk KOVÁŘ David SKLÁDAL Petr

Year of publication 2012
Type Article in Proceedings
Conference XII. Pracovní setkání fyzikálních chemiků a elektrochemiků
MU Faculty or unit

Faculty of Science

Citation
Web http://web2.mendelu.cz/af_239_nanotech/data/up/sbornik-mendelu.pdf#page=43
Field Biochemistry
Keywords Bacillus atrophaeus; Biosensor; QCM; SPR
Description Interactions of antibodies with the bacterium Bacillus atrophaeus were studied using piezoelectric (PZ) and surface plasmon resonance (SPR) based biosensors, with the aim of future development of biosensor for Bacillus species. Antibodies were covalently immobilised on gold surfaces of PZ crystal resonators and SPR chips. QCM Analyzer and Biacore 3000 were used for real-time and label-free measurements of immunocomplexes. Concentrations of B. atrophaeus below 10^5 CFU ml-1 were successfully detected.
Related projects:

You are running an old browser version. We recommend updating your browser to its latest version.

More info