Informace o publikaci

Microfluidic chip coupled with modified paramagnetic particles for sarcosine isolation in urine

Autoři

ZITKA Ondrej CERNEI Natalia HEGER Zbynek MATOUSEK Miroslav KOPEL Pavel KYNICKY Jindrich MASAŘÍK Michal KIZEK Rene ADAM Vojtech

Rok publikování 2013
Druh Článek v odborném periodiku
Časopis / Zdroj Electrophoresis
Fakulta / Pracoviště MU

Lékařská fakulta

Citace
Doi http://dx.doi.org/10.1002/elps.201300114
Obor Biochemie
Klíčová slova Ion exchange chromatography; Magnetic nanoparticles; Microfluidic separation; Prostate cancer; Surface modification
Popis Carcinoma of prostate (CaP) is the second most frequent malignant tumor occurring in men in Europe. Currently there is discussion on a wide range of potential CaP markers. One of them-nonprotein amino acid sarcosine, also known as N-methylglycine was chosen as a challenge for the development of microfluidic system with isolation by modified paramagnetic microparticles. Therefore, the aim of this study was to design a low-cost, simple, and rapid microfluidic system based on sarcosine isolation with modified paramagnetic microparticles and subsequent analysis on the ion exchange LC. We modified Dowex microparticles with Fe2O3 nanoparticles. Our paramagnetic microparticles were able to establish the binding with sarcosine. Moreover, we designed microfluidic device for sarcosine determination. Analysis of samples was carried out with LOD of 1 uM of a sarcosine that is sufficient because it is similar to concentrations of a sarcosine observed in the CaP patients.

Používáte starou verzi internetového prohlížeče. Doporučujeme aktualizovat Váš prohlížeč na nejnovější verzi.

Další info