Zde se nacházíte:
Informace o publikaci
Letter to the Editor in response to Posta et al. (2022): Pediatric morphometric study to guide the optimized implantation of the OSIA 2 implant system
Autoři | |
---|---|
Rok publikování | 2022 |
Druh | Článek v odborném periodiku |
Časopis / Zdroj | European Archives of Oto-Rhino-Laryngology |
Fakulta / Pracoviště MU | |
Citace | |
www | https://link.springer.com/article/10.1007/s00405-022-07593-3 |
Doi | http://dx.doi.org/10.1007/s00405-022-07593-3 |
Klíčová slova | Bone conduction implants; Osia; Children; Surgery |
Popis | The authors of this paper propose the OSIA BI300 bone conduction implant to be feasible for implantation in children (“non-complicated cases”), without preoperative CT scan. In my experience as a pediatric surgeon, there is great danger of hitting the dura or sigmoid sinus when blindly drilling into the bone without having first performed a CT scan (see LION surgery;) where the surgeon (Antonio Delle Volpe) drilled 4 holes in a 5 year-old till he found bone where the screw was stable). In fact, Posta et al. state that “For safety purposes, knowing the position of sigmoid sinus is also important.” |
Související projekty: |