Informace o publikaci

Automated Sequential Derivatization for Gas Chromatography-[Orbitrap] Mass Spectrometry-based Metabolite Profiling of Human Blood-based Samples

Logo poskytovatele
Autoři

JBEBLI Akrem COUFALÍKOVÁ Kateřina ZANABONI Moira BERGNA Manuela PICENONI Renzo KLÁNOVÁ Jana PRICE Elliott James

Rok publikování 2025
Druh Článek v odborném periodiku
Časopis / Zdroj Bio-protocol
Fakulta / Pracoviště MU

Přírodovědecká fakulta

Citace
www https://bio-protocol.org/en/bpdetail?id=5196&type=0
Doi http://dx.doi.org/10.21769/BioProtoc.5196
Klíčová slova Automation; Derivatization; Thermo ScientificTM TriPlusTM RSH; Metabolite profiling; Gas chromatography-mass spectrometry (GC-MS)
Přiložené soubory
Popis Many small molecules require derivatization to increase their volatility and to be amenable to gas chromatographic (GC) separation. Derivatization is usually time-consuming, and typical batch-wise procedures increase sample variability. Sequential automation of derivatization via robotic liquid handling enables the overlapping of sample preparation and analysis, maximizing time efficiency and minimizing variability. Herein, a protocol for the fully automated, two-stage derivatization of human blood-based samples in line with GC- [Orbitrap] mass spectrometry (MS)-based metabolomics is described. The protocol delivers a sample-to-sample runtime of 31 min, being suitable for better throughput routine metabolomic analysis.
Související projekty:

Používáte starou verzi internetového prohlížeče. Doporučujeme aktualizovat Váš prohlížeč na nejnovější verzi.

Další info