Informace o publikaci

Development of piezoelectric immunosensors for measurement of albuminuria

Autoři

NAVRÁTILOVÁ Iva SKLÁDAL Petr VIKLICKÝ Vladimír

Rok publikování 2001
Druh Článek v odborném periodiku
Časopis / Zdroj Talanta
Fakulta / Pracoviště MU

Přírodovědecká fakulta

Citace
Obor Biochemie
Klíčová slova Immunosensor; Quartz crystal microbalance; Human serum albumin; Albuminuria
Popis The development of piezoelectric immunosensors for human serum albumin (HSA) is reported. The piezoelectric crystals were modified either with monoclonal antibody AL-01 (direct assay) or with HSA (competitive assay). Measurements were carried out in the flow-through mode. Affinity interaction between albumin and the antibody was characterised. With immobilised antibody and HSA in solution, the kinetic association rate constant ka was 18100 l mol-1 s-1 and the dissociation constant kd was 0.00369 s-1. For the opposite arrangement (immobilised HSA), a slower dissociation was observed, kd was 0.00085 s-1. A competitive assay for HSA was developed with working range of 1 to 5 ug/ml and a total time for one analysis equal to 17 min. Samples of urine were analysed after tenfold dilution. The developed system was successfully evaluated on real samples from diabetic patients and the obtained results correlated well with the standard reflectometric assay of proteins in urine.
Související projekty:

Používáte starou verzi internetového prohlížeče. Doporučujeme aktualizovat Váš prohlížeč na nejnovější verzi.