You are here:
Publication details
Matrix interferences in determination of selenium by inductively coupled plasma - atomic emission spectrometry
Authors | |
---|---|
Year of publication | 2002 |
Type | Article in Proceedings |
Conference | 2002 Winter Conference on Plasma Spectrochemistry |
MU Faculty or unit | |
Citation | |
Field | Analytic chemistry |
Keywords | matrix interference; spectral interference; selenium; ICP-OES; signal enhancement |
Description | Spectral and non-spectral interferences observed at the determination of selenium by ICP-OES were investigated in this study. The identity of the carbon-induced emission peak that interferes on the Se 196.090 nm analytical line has been revised. It has been found that this peak is not a part of a CN band but it belongs most likely to a spectral band emitted by CO molecule. Non-spectral interferences of matrix elements C, S, P and Br consisting in the enhancement of selenium emission have been found. Concentration dependences of these matrix effects were measured in the range of 0.5 - 20 g.l-1 of matrix element and 0.1 - 100 mg.l-1 of Se. The enhancement of selenium emission was about 50% in the most significant case (sulfur 10 g.L-1). Similar effect was observed when gaseous compounds of C and S were introduced into the ICP. Mechanism of this matrix effect was studied and partially clarified as probably related to ionization of matrix elements. |
Related projects: |