Zde se nacházíte:
Informace o publikaci
Determination of Zr and Ti in 3-micrometers-thick ZrTiN ceramic coatings using laser ablation inductively coupled plasma atomic emission spectrometry
Autoři | |
---|---|
Rok publikování | 1997 |
Druh | Článek v odborném periodiku |
Časopis / Zdroj | Applied Spectroscopy |
Fakulta / Pracoviště MU | |
Citace | |
Obor | Analytická chemie, separace |
Klíčová slova | Atomic emission spectrometry; inductively coupled plasma; UV laser ablation; Coating analysis; Nitride ceramic coatings |
Popis | Analysis of 3-mu m-thick ZrN, TIN, and ZrTiN ceramic coatings on a high-speed steel substrate by laser ablation inductively coupled plasma atomic emission spectrometry (LA-ICP-AES) is described, A Nd:YAG laser was used in the Q-switched mode and operated at 355 nm and 5 mJ, Ablation patterns were obtained on the target surface by moving the ablation chamber with respect to the laser beam by means of an X-Y translator, Control of the depth of penetration was obtained by successively performing a limited number of cycles. Transient signals were obtained and their peak areas were used for the construction of calibration graphs based on the concentrations determined by energy-dispersive X-ray (EDX) analysis, An erosion rate of 0.1 mu m per cycle was observed, Crater shape and depth, temporal behavior, and reproducibility of the signals are described. |